Biochemistry Analyzer

The 2700 SELECT Biochemistry Analyzer provides vital information about your processes. It provides rapid, accurate analysis of key nutrients and byproducts, including:

- Glucose

- L-Lactate

- L-Glutamine

- L-Glutamate

- Ethanol

- Lactose

- Sucrose

- Galactose

- Hydrogen peroxide

- Methanol

- Choline

 

Sample analysis can be done two ways, off-line or on-line.

Off-Line Analysis

For many processes, off-line analysis is a good method for monitoring nutrients or byproducts, particularly if consumption or production rates are slow. As a benchtop Analyzer, the 2700 SELECT is easy to use and provides results quickly—in 60 seconds. Sample volume requirements are low (about 25 µL) which allow you to run as many samples as you want. And the best feature of all is that no sample preparation is required (no filtration and no dilution in most cases).

2710 Turntable

Couple the 2700 SELECT with the 2710 Turntable for batch analysis. The 24-position turntable lets you load samples and walk away. The instrument will automatically analyze them, print results and send them to a PC.

On-Line Analysis

For processes that require frequent monitoring and regulation, on-line analysis allows you the freedom to sample automatically and adjust both nutrient and byproduct concentrations.    

Couple the 2730 Aseptic Monitoring & Control system with the 2700 SELECT and pull samples directly from a bioreactor, measure important nutrients and byproducts, and immediately control the replenishment of nutrients, if needed, while maintaining the sterility of the process. All system components are sterilizable.

How On-Line Monitoring Works

The 2730 Monitoring and Control system pulls sample directly from a bioreactor and pumps it to the 2700 SELECT at programmed intervals. After each analysis, an antiseptic reagent, typically a dilute hypochlorite or hydroxide solution, is flushed through the sample lines and remains there until the next cycle to assure sterility.

For those bioprocesses in which loss of cells is a concern, you can connect the system to an on-line filtration device. You can adjust system parameters, such as flow rate, to accommodate the needs of the filtration device.

For regulation, connect the 2700 SELECT to a feed pump. When nutrient concentration falls below a programmed setpoint, the 2700 SELECT will signal the feed pump to add nutrient to the bioreactor.

Other system variables, such as sampling interval and concentration setpoint, allow you to customize the instrument for your process needs. For example, with processes that consume glucose quickly, you can configure the system to monitor glucose every few minutes. For other processes that consume nutrients slowly, monitoring every 30 or 60 minutes may be more
appropriate.

Other Features

RS232 and analog signal (with 2730) allow data logging with a PC

System can be set up for closed-loop monitoring and control

Discrete samples can be run when Analyzer is acting as on-line monitor

On-line monitor components can be autoclaved

Two measurement chemistries can be run simultaneously

Measurement units are selectable: g/L, mg/L, mmol/L, % (w/v)

Chemistries

Unless otherwise stated, specifications reflect use with YSI-supplied calibration solutions and a 25 µL sample.

Sample size, calibration concentration, and endpoints can be changed to accommodate specific applications.

Chemistry

Detection
Range (to)

Calibration
Point

Linearity Check Point

Precision
CV
n=10

Linearity
0 to cal point

Linearity
Cal point to range max

Typical Membrane Working Life

Choline

450 mg/L

175 mg/L

450 mg/L

2%

±2%

±5%

7 days

D-Glucose
Dextrose
(Low Range)

9 g/L
at 25 µL sample size

2.50 g/L

9.0 g/L
at 25 µL
sample size

2%

±2%

±5%

21 days

D-Glucose
Dextrose
(High Range)

25 g/L
at 10 µL sample size

2.50 g/L

25 g/L
at 10 µL
sample size

2%

±2%

±5%

21 days

Ethanol1

3.2 g/L

2.00 g/L

3.20 g/L

2%

±2%

±5%

5 days

Galactose2,3

25 g/L

N/A

N/A

N/A

N/A

N/A

10 days

Hydrogen
Peroxide2

600 mg/L

N/A

N/A

N/A

N/A

N/A

21 days

Lactose

25 g/L

5.00 g/L

25.0 g/L

2%

±2%

±5%

10 days

L-Glutamate

10 mmol/L

5.00 mmol/L

10.0 mmol/L

2%

±2%

±5%

7 days

L-Glutamine4

8 mmol/L

5.00 mmol/L

8.00 mmol/L

4%

±4%

±5%

5 days

L-Lactate
L-Lactic Acid

2.67 g/L

0.50 g/L

2.67 g/L

2%

±2%

±5%

14 days

Methanol5

2.50 g/L

1.00 g/L

2.50 g/L

2%

±2%

±5%

5 days

Sucrose

25 g/L

5.00 g/L

25.0 g/L

2%

±2%

±5%

10 days

 

1. Methanol is a potential substrate interference. Specifications are for 10 µL sample.
2. The user must provide calibrator solution.
3. Lactose is a potential substrate interference.
4. Specifications are for the simultaneous measurement of glutamine and glutamate.
5. Ethanol is a potential substrate interference. Specifications are for 15 µL sample.

 



Liquid analysis/Biochemistry Analyser
                 Request for information
Company
Name
Country
Phone
Email
Your question
Verification letters are small [0 is a zero] and [o is a letter O]

Copyright: Ankersmid 2010 Contact: info@ankersmidmci.com